The Additel ADT672 series digital pressure calibrators with HART communication capability – Pressures to 60000 psi (4200 bar). Pressure Types: Gauge, Compound Absolute and Differential. 11 Selectable Pressure Units.
Additel Gauge Resolution Data Sheet
At first glance, the Additel ADT672 series precision pressure calibrators look like an ordinary pressure gauge. But this series is much more than ordinary, and definitely more than just a pressure gauge—it’s a pressure calibrator!
The ADT672 has advanced microprocessor technology and state-of-the-art silicon pressure sensors, the ADT672 series precision pressure calibrators provide a pressure calibration solution for gauges, transmitters, and switches over a wide pressure range. The ADT672 is the size of a pressure gauge but with the functionality of a calibrator: It measures pressure precisely with a builtin pressure sensor, as well as reads the current or mV produced by a transducer. It can even supply an excitation voltage to power sensors or transmitters during calibration. In order to reach 0.02%FS accuracy up to 10,000 psi (700 bar) and 0.1%FS accuracy up to 60,000 psi (4,200 bar), every silicon pressure sensor has been specially aged, tested, and screened before assembly. The 672 series precision pressure calibrators are unmatched in performance and reliability.
- Model: ADT672
- Pressure ranges to 60,000 psi (4,200 bar)
- HART Communication capability
- Measure mA or V, and with 24V loop power
- Easy-to-use, inexpensive pressure calibrator with uncertainty better than 0.02%FS
- ISO 17025-accredited calibration w/data included
Applications:
- Gauge calibration
- Precision pressure measurement
- Pressure transmitters calibration
- Pressure switch testing
- Safety relief valve resting
- Pressure regulator testing
- Hydraulic leak testing
MAXIMATOR specialises in supporting your business further by offering the following:
- Local Technical support
- NATA Test Certificates
- Customised Packages
Videos:
Additel 672 Video Demo
DP Gauge Calibration Video Demo
Pressure Switch Video Demo
- Pressure ranges to 60,000 psi (4,200 bar)
- Measure mA with 0.01% RD + 1.5 μA accuracy Measure V with 0.01% RD + 1.5 mV accuracy
- Power transmitters during test using 24V loop supply
- Pressure switch test HART Communication capability
- Advanced temperature compensation
- Dual readout Min/Max/Hold to capture changing measurements
- Data logging
- Large, easy to read display with 6-digit resolution
- Backlit display
- Rechargeable battery or AC adapter
- ISO17025 accredited calibration with data (Included)
Gauge Pressure[1] | |||||
---|---|---|---|---|---|
P/N | Pressure Range | Media[2] | Accuracy(%FS) | Burst Pressure | |
psi | bar | ||||
V15 | -15 | -1 | G | 0.02 (0.05) | 3× |
GP2 | 2 | 0.16 | G | 0.05 | 3× |
GP5 | 5 | 0.35 | G | 0.05 | 3× |
GP10 | 10 | 0.7 | G | 0.02 (0.05) | 3× |
GP15 | 15 | 1 | G, L[3] | 0.02 (0.05) | 3× |
GP30 | 30 | 2 | G, L[3] | 0.02 (0.05) | 3× |
GP50 | 50 | 3.5 | G, L | 0.02 (0.05) | 3× |
GP100 | 100 | 7 | G, L | 0.02 (0.05) | 3× |
GP150 | 150 | 10 | G, L | 0.02 (0.05) | 3× |
GP300 | 300 | 20 | G, L | 0.02 (0.05) | 3× |
GP500 | 500 | 35 | G, L | 0.02 (0.05) | 3× |
GP600 | 600 | 40 | G, L | 0.02 (0.05) | 3× |
GP1K | 1000 | 70 | G, L | 0.02 (0.05) | 3× |
GP2K | 2000 | 140 | G, L | 0.02 (0.05) | 3× |
GP3K | 3000 | 200 | G, L | 0.02 (0.05) | 3× |
GP5K | 5000 | 350 | G, L | 0.02 (0.05) | 3× |
GP10K | 10000 | 700 | G, L | 0.02 (0.05) | 3× |
GP15K | 15000 | 1000 | G, L | 0.05 (0.1) | 2× |
GP20K | 20000 | 1400 | G, L | 0.05 (0.1) | 1.5× |
GP25K | 25000 | 1600 | G, L | 0.1 | 1.5× |
GP30K | 30000 | 2000 | G, L | 0.1 | 1.5× |
GP36K | 36000 | 2500 | G, L | 0.1 | 1.5× |
GP40K | 40000 | 2800 | G, L | 0.1 | 1.35× |
GP50K | 50000 | 3500 | G, L | 0.1 | 1.2× |
GP60K | 60000 | 4200 | G, L | 0.1 | 1.1× |
[1] Sealed gauge pressure for above 1,000 psi [2] G=Gas, L=Liquid [3] 0.02% FS for gas media only |
Compound Pressure | |||||
---|---|---|---|---|---|
P/N | Pressure Range | Media | Accuracy(%FS)[1] | Burst Pressure | |
psi | bar | ||||
CP2 | ±2 | ±0.16 | G | 0.05 | 3× |
CP5 | ±5 | ±0.35 | G | 0.02 (0.05) | 3× |
CP10 | ±10 | ±0.7 | G | 0.02 (0.05) | 3× |
CP15 | ±15 | ±1 | G | 0.02 (0.05) | 3× |
CP30 | -15 to 30 | -1 to 2 | G | 0.02 (0.05) | 3× |
CP100 | -15 to 100 | -1 to 7 | G, L | 0.02 (0.05) | 3× |
CP300 | -15 to 300 | -1 to 20 | G, L | 0.02 (0.05) | 3× |
[1] FS specification applies to the span of the range |
Absolute Pressure | |||||
---|---|---|---|---|---|
P/N | Pressure Range | Media | Accuracy(%FS) | Burst Pressure | |
psi | bar | ||||
AP5 | 5 | 0.35 | G | 0.1 | 3× |
AP10 | 10 | 0.7 | G | 0.1 | 3× |
AP15 | 15 | 1 | G | 0.1 | 3× |
AP30 | 30 | 2 | G | 0.1 | 3× |
AP50 | 50 | 3.5 | G | 0.1 | 3× |
AP100 | 100 | 7 | G , L | 0.05 (0.1) | 3× |
AP300 | 300 | 20 | G , L | 0.05 (0.1) | 3× |
AP500 | 500 | 35 | G , L | 0.05 (0.1) | 3× |
AP1K | 1000 | 70 | G , L | 0.05 (0.1) | 3× |
AP3K | 3000 | 200 | G , L | 0.05 (0.1) | 3× |
AP5K | 5000 | 350 | G , L | 0.05 (0.1) | 3× |
Differential Pressure | ||||||
---|---|---|---|---|---|---|
P/N | Pressure Range | Media | Accuracy(%FS)[1] | Burst Pressure | Static Pressure Range | |
inH2O | mbar | |||||
DP1 | ±1 | ±2.5 | G | 0.05[2] | 100× | ±10 psi |
DP2 | ±2 | ±5.0 | G | 0.05[2] | 100× | ±10 psi |
DP5 | ±5 | ±10 | G | 0.05[2] | 50× | ±10 psi |
DP10 | ±10 | ±25 | G | 0.05[2] | 20× | ±10 psi |
DP20 | ±20 | ±50 | G | 0.05 | 20× | ±10 psi |
DP30 | ±30 | ±75 | G | 0.05 | 20× | ±10 psi |
DP50 | ±50 | ±160 | G | 0.05 | 3× | ±10 psi |
DP150 | ±150 | ±350 | G | 0.02 | 3× | 50 psi |
DP300 | ±300 | ±700 | G | 0.02 | 3× | 50 psi |
[1] FS specification applies to the span of the range. Accuracy includes one year stability. [2] 0.05%FS accuracy (incl 6 months stability). One year accuracy is 0.05%FS calibration accuracy combined with 0.05%FS one year stability. |
Specification | |
---|---|
Accuracy | ADT672-02: 0.02% of full span |
ADT672-05: 0.05% of full span | |
>20,000 (1,400 bar): 0.1% FS | |
Gauge Types | Gauge pressure, compound pressure, absolute pressure, differential pressure |
Display | Description: Dual-line 6 full digit FSTN LCD with LED Backlight |
Display rate: 3.5 readings per second (Default setting). | |
Numeral display height: 16.5mm (0.65″) | |
Pressure Units | Pa, kPa, MPa, psi, bar, mbar, kgf/cm2, inH2O@4°C mmH2O@4°C, inHg@0°C, mmHg@0°C |
Environmental | Compensated Temperature: 14°F to 122°F (-10°C to 50°C) |
Operating Temperature: 14°F to 122°F (-10°C to 50°C) | |
Storage Temperature: -4°F to 158°F (-20°C to 70°C) | |
Humidity:<95% | |
Pressure Port | ≤15,000 psi: 1/4NPT male, 1/2NPT male, 1/4BSP male, 1/2BSP male, M20×1.5 male |
>15,000 psi: 1/4HP female or 1/4HP male *1/4HP female: Autoclave F-250-C, 9/16″ – 18 UNF-2B *1/4HP male: Autoclave M-250-C, 9/16″ – 18 UNF-2A | |
Differential Pressure: 0.236 inch (Ø6 mm) test hose | |
Other connections available per request | |
Over Pressure Warning | 1.2 |
Electrical Connection | 0.156 inch (Ø4mm) sockets |
Electrical Measurement Accuracy | Voltage DC: ±30.0000V, ± (0.01%RD + 1.5 mV) |
Current DC: ±30.0000mA, ± (0.01%RD + 1.5 μA) | |
DC 24V: 24V±0.5V, MAX:50mA, Protect at: 120mA | |
Switch[1]: Status OPEN/CLOSED | |
Power | Battery: Rechargeable Li-ion polymer battery |
Li-Battery working time: 40 hours | |
Recharge time: 4 hours | |
External power: 110V/220V power adapter (DC10V) | |
Enclosure | Case material: Aluminum alloy |
Wetted parts: 316L SS | |
Dimension: Ø120mm X 46mm depth X 184mm height | |
Weight: 0.7kg | |
Data Logging | Storage capacity: 30 files, 40 records per file |
Mode: manual and automatic | |
Hourly-record: record the data every hour | |
Interval-record: set by user | |
Compliance | CE Marked |
Communication | RS232 (DB9/F, environmentally sealed) |
Baud rate: 1200, 2400, 4800, 9600 | |
Data length: 8 bits | |
Stop bit: 2 bits | |
Address: from 1 to 112 | |
Warranty | 1 year |